HC10900 — Vacuum pump motor M13 not connected or malfunctioning
Agfa CT error code, reported by the Film Pick-up Module.
What HC10900 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10900 indicates: Vacuum pump motor M13 not connected or malfunctioning. The code originates in the Film Pick-up Module — Vacuum Pump (M13) section of the Film Pick-up Module. Severity level alarm (Alarm).
Reported action: Check the connections to motor M13, then connector XS4 and connector P31 on the MEC board — reconnect or replace any connector that isn't OK. If the connections check out, verify the wiring from M13 to MEC/P31 with a multimeter and fix it if faulty. If the wiring is good, power off the printer and replace M13; if the fault persists, the MEC board itself is defective and needs replacing..
Raised on the Drystar 5500/5503 when vacuum pump motor M13 in the film pick-up module is not connected or is malfunctioning. Filed in the appendix's pick-up module fault-code table alongside the tray's other sensor, motor, magazine-lock and tag-reader codes (HC10803-HC10813, HC10818, HC10821-HC10822, HC10836, HC10906, HC10916-HC10918, HC10921-HC10924, HC11018-HC11019, HC11021).
Full troubleshooting for HC10900
The complete diagnosis, cause tree, and step-by-step remedy for the Film Pick-up Module — Vacuum Pump (M13) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Film Pick-up Module — Vacuum Pump (M13) codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.