HC10824 — Alignment module vertical home-position sensor S5 not connected or malfunctioning
Agfa CT error code, reported by the Alignment Module.
What HC10824 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10824 indicates: Alignment module vertical home-position sensor S5 not connected or malfunctioning. The code originates in the Alignment Module — Home Position Sensors (S5/S6) section of the Alignment Module. Severity level alarm (Alarm).
Reported action: Check the connections to sensor S5, then connector P27 on the MEC board — reconnect or replace whichever isn't OK. If the connections are fine, check the wiring from S5 to MEC/P27 with a multimeter and fix it if faulty. If the wiring is good, power off the printer and replace S5; if the fault persists, the MEC board is defective and needs replacing..
Raised on the Drystar 5500/5503 when the alignment module's vertical home-position sensor S5 is not connected or is malfunctioning. The printer shows "Call service". Filed alongside the equivalent horizontal home-position code, HC10825, and the homing time-out code HC11208 that also involves S5.
Full troubleshooting for HC10824
The complete diagnosis, cause tree, and step-by-step remedy for the Alignment Module — Home Position Sensors (S5/S6) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Alignment Module — Home Position Sensors (S5/S6) codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.