AlarmDrum Module — THIB Ambient Temperature Sensor

HC10817Ambient temperature out of window at the drum module (THIB)

Agfa CT error code, reported by the Print Drum Module.

What HC10817 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10817 indicates: Ambient temperature out of window at the drum module (THIB). The code originates in the Drum Module — THIB Ambient Temperature Sensor section of the Print Drum Module. Severity level alarm (Alarm).

Reported action: Two documented causes: site temperature outside the printer's installation specification, or a malfunctioning ambient-temperature measurement circuit on the THIB. Confirm site temperature against the installation specification first; if it's within spec and the code persists, power off the printer and replace the THIB board..

Raised on the Drystar 5500/5503 when the ambient-temperature reading taken on the THIB (Thermal Head Interface Board), inside the drum module, falls outside the printer's allowed window. Classed as a Defect — the printer shows "Ambient temperature problem" and will not print until it's resolved. Its counterparts HC11227/HC11228 measure ambient temperature on the MEC board instead and are only Accidental-level.

Full troubleshooting for HC10817

The complete diagnosis, cause tree, and step-by-step remedy for the Drum Module — THIB Ambient Temperature Sensor subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Drum Module — THIB Ambient Temperature Sensor codes

Discuss HC10817 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.