AlarmDrum Module — THIB Ambient Temperature Sensor

HC10817 — Ambient temperature out of window at the drum module (THIB)

Agfa CT error code, reported by the Print Drum Module.

What HC10817 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10817 indicates: Ambient temperature out of window at the drum module (THIB). The code originates in the Drum Module — THIB Ambient Temperature Sensor section of the Print Drum Module. Severity level alarm (Alarm).

Reported action: Two documented causes: site temperature outside the printer's installation specification, or a malfunctioning ambient-temperature measurement circuit on the THIB. Confirm site temperature against the installation specification first; if it's within spec and the code persists, power off the printer and replace the THIB board..

Raised on the Drystar 5500/5503 when the ambient-temperature reading taken on the THIB (Thermal Head Interface Board), inside the drum module, falls outside the printer's allowed window. Classed as a Defect — the printer shows "Ambient temperature problem" and will not print until it's resolved. Its counterparts HC11227/HC11228 measure ambient temperature on the MEC board instead and are only Accidental-level.

Full troubleshooting for HC10817

The complete diagnosis, cause tree, and step-by-step remedy for the Drum Module — THIB Ambient Temperature Sensor subsystem are documented in the FSELIB Agfa service manual — re-authored, web-first, with real figures.

7-day free trial · no credit card required

Related Drum Module — THIB Ambient Temperature Sensor codes

Discuss HC10817 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.

“That they may see, and know, and consider, and understand together, that the hand of the LORD hath done this.”Isaiah 41:20