HC10802 — Jam-clearance toggle switch SW5 not connected or malfunctioning
Agfa CT error code, reported by the Cover-Open & Jam-Clearance Switches.
What HC10802 means
On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10802 indicates: Jam-clearance toggle switch SW5 not connected or malfunctioning. The code originates in the Cover-Open & Jam-Clearance Switches section of the Cover-Open & Jam-Clearance Switches. Severity level alarm (Alarm).
Reported action: Check the connections to switch SW5, then connector P9 on the MEC board — reconnect or replace whichever isn't OK. If the connections are fine, check the wiring from SW5 to MEC/P9 with a multimeter and fix it if faulty. If the wiring is good, power off the printer and replace SW5; if the fault persists, the MEC board is defective and needs replacing..
Raised on the Drystar 5500/5503 when the jam-clearance toggle switch SW5 is not connected or is malfunctioning. Filed with the other cover-open and jam-clearance switch codes, HC10800 (SW1) and HC10801 (SW2).
Full troubleshooting for HC10802
The complete diagnosis, cause tree, and step-by-step remedy for the Cover-Open & Jam-Clearance Switches subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.
Open the Activion16 service manualRelated Cover-Open & Jam-Clearance Switches codes
Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.