Pular para o conteúdo principal
AlarmCover-Open & Jam-Clearance Switches

HC10800Cover-open switch SW1 not connected or malfunctioning

Agfa CT error code, reported by the Cover-Open & Jam-Clearance Switches.

What HC10800 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC10800 indicates: Cover-open switch SW1 not connected or malfunctioning. The code originates in the Cover-Open & Jam-Clearance Switches section of the Cover-Open & Jam-Clearance Switches. Severity level alarm (Alarm).

Reported action: Check the connections to switch SW1, then connector P9 on the MEC board — reconnect or replace whichever isn't OK. If the connections are fine, check the wiring from SW1 to MEC/P9 with a multimeter and fix it if faulty. If the wiring is good, power off the printer and replace SW1; if the fault persists, the MEC board is defective and needs replacing..

Raised on the Drystar 5500/5503 when cover-open switch SW1 is not connected or is malfunctioning. Filed with the other cover-open and jam-clearance switch codes, HC10801 (SW2) and HC10802 (SW5).

Full troubleshooting for HC10800

The complete diagnosis, cause tree, and step-by-step remedy for the Cover-Open & Jam-Clearance Switches subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Cover-Open & Jam-Clearance Switches codes

Discuss HC10800 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.