Phoenix PDU Main Breaker Trips, Will Not Reset due to Failed TVSS Module.

Symptom

Phoenix PDU Main Breaker Trips, Will Not Reset

Applies to

Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Component
2185383-11 MOV MODULE Also known as "TVSS". When FE's attempt to order the TVSS module listed in HFO3 Renewal Parts: 2228875-7, FEMC automatically "changes up" to p/n 2185383-11 MOV MODULE instead, because 2228875-7 cannot be obtained anymore. Phoenix is out of business. I checked with Margie Wiza of MR Eng and this is as intended. This is to fix PQR # 13109044.
Area
MR

Full cause & step-by-step resolution