The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.

Symptom

The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered See Long Description The carriage fails to advance to meet the table when it is raised, and fails to retract when it is undocked or lowered. But the customer can manually attach the cradle to the carriage and then table operation is normal driving in and out and they can scan okay. There are no error messages logged.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
46-271465P1 SENSOR FLAG MOUNTING CLAMP BLOCK 46-265486P8 SCREW Slotted Pan Head Delrin (Natural White plastic) Machine Screw 6-32 X 0.31 Slotted Pan Head Delrin (Natural White plastic) Machine Screw 6-32 X 0.31
Area
MR