GE Healthcare · MRExcite225271222493381081934956106392000023562249352
3T system has intermittent scan stops and screen lockups which require TPS reset to clear. 3T Excite G3.
Symptom
Scan Stops Intermittently and the Screen Locks Up
Error message
2252712 Sequence playout error. EOS from one or more sequence generators was not received prior to SSP EOS. Wed Sep 1 19:14:27 2004 Error: 2252712 Host: MRI1-OC0 Process: NSP File: AGP:agp_drv.c Line: 411 Sequence playout error. EOS from one or more sequence generators was not received prior to SSP EOS. Generator: Theta 2 2249338 CM init failed. msgQSmCreate failed when creating VxMP mailbox. Error seen with MGD Diags: 1081934956 4 1 Wed Apr 14 09:29:16 2004 2249338 2 CARD diagCoord cm_mp.c 306 Task ID: 0x3a652d0 Time: 10639 Object: APS CM init failed. msgQSmCreate failed when creating VxMP mailbox. errno = 0x3D0001.-> 200002356 dsserver died unexpectedly, no action being taken. Mon Nov 22 10:46:20 2004 Error: 200002356 Host: MRI1-OC0 Process: hostMonitor File: ssProcStop.c Line: 906 dsserver died unexpectedly, no action being taken. 2249352 Exception at interrupt level: program See long description. Wed Oct 6 08:09:53 2004 Error: 2249352 Host: MRI1-OC0 Process: diagCoord File: bootConfig.c Line: 3639 Task ID: 0x3ee8d08 Time: 40 Object: APS Exception at interrupt level: program Exception current instruction address: 0x02a63210 Machine Status Register: 0x00089030 Condition Regist
Applies to
- Models
- Excite
- Product
- 3T Excite G3 Software ONLY
- Software
- G3 For 3T Excite only
- Component
- 2294300-4 APS BOARD WITH IT DAUGHTER BD
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.