A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During S…

A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.

Symptom

Coolant Leak at the MCS Cabinet Chiller. See Long Description When the FE powered on the HDe MCS Cabinet Chiller for the first time, all of the coolant leaked out.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
5165290 MCS FRU KIT HDe MCS FRU KIT HDe 5165290 MCS FRU KIT HDe 5165290
Area
MR