Pular para o conteúdo principal
GE Healthcare · MRExcite2220017

restoreinfo Fails Due to Improper Coil Config Update.

Symptom

Coil Config File is Missing after 12X or Earlier Upgrade to 14.0 M5 After RestoreINFO and during load of 14.0 M5 Apps software a popup error message says CoilConfig.cf.1.5.T file does not exist. Click on OK to exit the tool. This message will NOT appear for 3.0T systems, even though the problem MAY exist. The steps shown in the linked Resolution should be performed for all 3.0T upgrades as well.

Error message

CoilConfig.cfg.1.5T file does not exist. Click on OK to exit the tool. Pop up message happens after RestoreINFO and during load of 14.0 M5 Apps

2220017 Coil may not be listed in the proper group on Coil dialog box. Caused by CoilConfig File Error, during upgrade. Tue Mar 18 14:41:45 2008 Error: 2220017 Host: lx_wo Process: scn File: SrxCoilName.m Line: 2650 userInterfaceGroup field doesn't exist, is empty or is out of range for HEAD coil. Coil may not be listed in the proper group on Coil dialog box.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required