restoreinfo Fails Due to Improper Coil Config Update.
Symptom
Coil Config File is Missing after 12X or Earlier Upgrade to 14.0 M5 After RestoreINFO and during load of 14.0 M5 Apps software a popup error message says CoilConfig.cf.1.5.T file does not exist. Click on OK to exit the tool. This message will NOT appear for 3.0T systems, even though the problem MAY exist. The steps shown in the linked Resolution should be performed for all 3.0T upgrades as well.
Error message
CoilConfig.cfg.1.5T file does not exist. Click on OK to exit the tool. Pop up message happens after RestoreINFO and during load of 14.0 M5 Apps 2220017 Coil may not be listed in the proper group on Coil dialog box. Caused by CoilConfig File Error, during upgrade. Tue Mar 18 14:41:45 2008 Error: 2220017 Host: lx_wo Process: scn File: SrxCoilName.m Line: 2650 userInterfaceGroup field doesn't exist, is empty or is out of range for HEAD coil. Coil may not be listed in the proper group on Coil dialog box.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.