ID99497 Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.

Symptom

Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76

Error message

3602019 Bad kV Detected DO XRTechMgr acquisition Component(s) _Tube 1 0 [XRTechMgr::notifyGeneratorFinalExposureParameters] numberOfTubeSpits=X above the maxNumberOfTubeSpits=X after X frames XRTechMgr.cpp 1964 3602019 Bad kV Detected DO XRTechMgr acquisition Component(s) _Tube 1 0 [XRTechMgr::notifyGeneratorFinalExposureParameters] numberOfTubeSpits=X above the maxNumberOfTubeSpits=X after X frames XRTechMgr.cpp 1964

INX100: 3602069 Jedi: Spit or kV instability warning JEDI Error Code: 0324H Tube spit - spit cathode (yes) 3602069 Jedi: Spit or kV instability warning JEDI Error Code: 0324H Tube spit (all kV drop/regul errors) Spit Info: spit anode (no) - spit cathode (yes) - spit both (no) - kV_regul (no) Tube 1 70C: OK, Tube 40C: OK, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK, Door Status:OK

INX100: 3602036 Jedi: Power module problem JEDI Error Code: 0313H ILR max current detected 3602036 Jedi: Power module problem AR XRTechMgr acquisition JEDI_Power_Module_Problem 1 0 JEDI Error Code: 0313H ILR max current detected Data Related to Error: 0 Tube 1 70C: OK, Tube 40C: Set, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK,

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Innova Jedi Innova Jedi Innova Jedi
Area
X-ray Vascular