INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency…
INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
Symptom
INX1X1 - LP4 lift Limit switch engaged, Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit switch. Error message, Limit switch engaged; please call Service. (Error 1050458). Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because lift hit Max SID limit switch. Image display error = Limit switch engaged; please call Service. See LD for more details. Lateral Detector lift will not move after using the Emergency Backout button on the lift cover to drive it to maximum SID. Attempts to drive the lift from the TSUI (SH - Smart Handle or SB - Smart Box) do not work. Image display error = Limit switch engaged; please call Service. (Error 1050458)
Error message
Error: 1050458 - Lat: LAT DET LIFT AXIS SWITCH ENGAGED Debug: Limit switch 2 problem (prd: BC) Error: 1050458 - Lat: LAT DET LIFT AXIS SWITCH ENGAGED - Limit switch 2 problem (prd: BC) See LD for Error details, Root Cause, and possible Recommended Actions. Error Code: 1050458 Subsystem: POS Error Text: Lat: LAT DET LIFT AXIS SWITCH ENGAGED Recovery Class: WR (WaRning) Debug Text: Limit switch 2 problem (prd: BC) Root Cause: POS: Det Lift axis limit switch error for lateral plane. It is related to a problem detected on the Det Lift axis limit switch for lateral plane.
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Software
- Innova Biplane SPAD_M3 Innova Biplane SPAD_M3 Innova Biplane SPAD_M3
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.
- DL system hangs while customer trying to change protocols