Center Frequency is Missing a Digit Due to Being Typed Incorrectly During a Manual Prescan.

Symptom

MRIQ FAT SAT Is Not Working At All Or Is Uneven

Error message

2248265 The system tried to set the narrowband Exciter2 to XXXX Hz. The narrowband exciter can not handle this frequency. Mon Dec 13 16:36:07 2004 Error: 2248265 Host: MR05 Process: NSP File: APS:cex_set_center_freq Line: 1246 The system tried to set the narrowband Exciter2 to 113924133 Hz. The narrowband exciter can not handle this frequency. ExciterId = 0 (0=Master, 1=Slave, 2=both)

2224567 Auto Prescan incomplete. Could not identify fat and water peaks. Set Center Frequency manually. Fri Sep 24 09:41:26 2004 2224567 LHMR PscMainControl.cc 1172 Auto Prescan incomplete. R1=13 R2=15 TG=138. Could not identify fat and water peaks. Set Center Frequency manually.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.