Pular para o conteúdo principal
GE Healthcare · MRExcite22540122247157

Unable to Download or Reset The TPS MGD or ISE.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2254012 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. Rx Signal Detect signal is in an Invalid State. Tue May 14 19:20:37 2002 2254012 1 bay12b_oc0 NSP SCP:RRF-DIF 630 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. Rx Signal Detect signal is in an Invalid State.

2247157 The Master Exciter in the RRF chassis (rear, slot 7) was not detected. Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board.-> Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board. Extended Error: The master exciter provides the main clock for the RRF chassis and the IRF board. Verify that the board is powered and that its LO DETECT LED is on. If this is a 16 channel system, verify that the two fiber cables on the two DIF boards are not switched with each other. Possible cause(s): IRF Board, Exciter Board, RRF-DIF Board, GigaLink Fiber Cable

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.