Pular para o conteúdo principal
GE Healthcare · MRExciteSigna22238780016311194621391

Application Software does not complete the boot process with TPS Download failure Due to ICN Image Compute Node.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2223878 A process has crashed on the VRE (00 30 48 59 F0 64[icn1]). Process dpserver, signal causing the crash 7, time 001631, date: 2007-11-10, UTC 1194621391 2223878 HDe_mfg NSP VREReset.cc 1008 Exception Class: Unknown Function: Recon: Data Acquisition A process has crashed on the VRE (00:30:48:59:F0:64[icn1]). Process: dpserver, signal causing the crash: 7, time: 00:16:31, date: 2007-11-10, UTC: 1194621391. Extended Error: During a TPS Reset the VRE gets scanned for core files that were generated since the last TPS Reset or System Restart. Only one such message is generated even if there is more than once core file. The oldest core file found across all ICNs is listed in this message.

Applies to

Models
Excite, Signa
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues