GE Healthcare · X-ray VascularInnova3800331390033136020853602016
Pixel Gain Calibration Fails to complete
Symptom
INX100: Pixel Gain Calibration Fails to complete Pixel Gain Calibration Fails to complete. Any defect which causes the Pixel Gain calibration to fail to report success at the end of calibration.
Error message
3800331 or 3900331 for Lat. Bad actual techniques received in acquisition DO XRTechOpt EM Component(s)_Tube 1 1 (null) XRPAtlasLogger.cppXXX 3800331 or 3900331 for Lat.1 Bad actual techniques received in acquisition DO XRTechOpt EM Component(s)_Tube 1 1 (null) XRPAtlasLogger.cppXXX The Jedi has detected bad actual techniques during the acquisition, it will restart Xray. If the system issue continues, error 3800331 will be logged 6 times and xray aborted on the 7th fault. If X-ray is aborted review the error code imediately after the 6 3800331 errors logged in the log for the root cause of the exposure abort. KV REGULATION SPITS. 3602085, error Text = Jedi: too many tube spits, debug Text += jedi-error=316h, Spit Max error, data related to error: 8, recovery Class = AR, major Function /= XRTechMgr KV REGULATION SPITS. System impact: X-ray unavailable. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service". May be preceded by error 3800331 Bad actual techniques recieved in aquisition, and followed by 3602016 Degraded Image Quality
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.