GE Healthcare · X-ray VascularInnova50110
When driving the Gantry ARC to steep angles the Gantry is Stopping and Freezing.
Symptom
LC: When driving the Gantry ARC to steep angles the Gantry is Stopping and Freezing. When driving the Gantry arc to steep angles the Gantry is Stopping and Freezing.
Error message
50110 POS : CARC OVER CURRENT ERROR 4ABB HDW 0002 USR = 0200 50110 POS SMA: Over current error DO LC Gantry (null) Component(s) Positioner SMA Mechanical 1 0 POS ERROR=4ABB,HDW STATUS=0002,USR STATUS=0200 XRPAtlasLogger.cpp 4ABB HW 0002 User 0200 => Frontal arc overcurrent Note : HW error / Axis relationship 0x0000 HW_FRT_L 0x0001 HW_FRT_PVT 0x0002 HW_FRT_ARC 0x0003 HW_FRT_II User error signification 0x0100 US_AXIS_HARD_OVER_CURRENT 0x0200 US_AXIS_FAST_OVER_CURRENT 0x0400 US_AXIS_SLOW_OVER_CURRENT
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- Positioner Gantry Positioner Gantry Positioner Gantry
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.