No Table Position Display due to Cable Connector Fell Off J12 of the Table Electronics Board Internal to the Table.

Symptom

No Table Position Display but Table Moves In/Out and Up/Down See Long Description The system is an OpenSpeed. The table moves up/down and in/out and the alignment light comes on, but there is never any table position display on the magnet enclosure. It is always dark. The customer has already cycled all appropriate breakers and powered down the complete system, no help.

Applies to

Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Area
MR