GE Healthcare · MRExcite2254052

TG reached high limit on Head Coil due to cable/quick disconnect box.

Symptom

Auto Prescan Fails with TG Reached the High Limit See error in Long Description 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan.

Error message

2254052 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and Patient Weight Fri Aug 20 08:54:29 2004 Error: 2254052 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2384268 CLASSIC HEAD COIL CBL ASM W/COIL ID See Long Description and Attachment See Attachment for the engineering drawing of this part: 2384268ADW Rev. 2. This is not available online. To confirm this is the most appropriate drawing, go to ePDM.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.