OrthoMRI Power cable loose on gradient amplifier.OrthoMRI Scan is disabled.OrthoMRI Reseated and tie strapped the gradie…
OrthoMRI Power cable loose on gradient amplifier.OrthoMRI Scan is disabled.OrthoMRI Reseated and tie strapped the gradient amplifier power cable.OrthoMRI Escan (Unix Digital).OrthoMRI Error code 1901…
Symptom
OrthoMRI Scan is disabled
Error message
OrthoMRI Error code 1901 gradient amplifier not powered. Scan disabled
Applies to
- Product
- OrthoMRI Escan (Unix Digital) Two easy ways to tell if the Escan Unix system is analog or digital. First, the analog service password is coil and the digital password is probe. Second, the analog daily QA phantom is a bottle where as the digital daily QA phantom is a large spherical ball.
- Software
- OrthoMRI Unix 2.x
- Area
- MR
Full cause & step-by-step resolution