98438 Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.
Symptom
Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76
Error message
INX100: 3602038 IH XRTechMgr Jedi: kvctl configuration problem JEDI Error Code: 0701H Saved RAM checksum pb 3602038 IH XRTechMgr Jedi: kvctl configuration problem Debug: jedi-error-code=701h JEDI Error Code: 0701H Saved RAM checksum pb Data Related to Error: 0 Tube 1 70C: OK, Tube 40C: OK, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK, Door Status: Closed Phase: STAND_BY Generator PPC kV control board has detected corruption in the NV Ram verification (After power up) INX100: 3602075 Jedi: Unkwnon Inhibit error; Generic error ''Generic_class4'' reported for JEDI Error Code: 0557H 3602075 Jedi: Unkwnon Inhibit error IH XRTechMgr acquisition Jedi_Unknown_IH_error 0 0 JEDI Error Code: 0557H Generic error ''Generic_class4'' reported for JEDI Error Code: 0557H Data Related to Error: 144 Tube 1 70C: OK, Tube 40C: Set, Rotor Thermal Error: OK, Inverter thermal Error: OK, Tank thermal Error: OK, Door Status: Closed Phase: POW XRPAtlasLogger.cpp 156 Jedi Error 557H - Root Cause: LVPS TRI: Detected +160V too low +160V is lower than 110V (measured on heater board). 3602075 Jedi: Unkwnon Inhibit error IH XRTechMgr acquisition Jedi_Unknown_IH_error 0 0 JEDI Error
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- Innova Jedi Innova Jedi Innova Jedi
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.