Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedVCT260107214260107212

BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4

Symptom

Intermittent rotated images - may have no errors in the log. See Long Description for details Intermittent rotated images - may have no errors in the log.. Rotated images in all of the helical series. Rotated Image two.

Error message

Error: 260107214 Gantry Failed to Stop at Position XX degrees within XX.X seconds See Long Description for complete error Thu Sep 22 13:19:47 2005 Error: 260107214 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3687 Thu Sep 22 13:19:47 2005 Host : Table Gantry Processor Ermes # : 260107214 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3687 Function : AXIAL CONTROL : Axial Control Scan Type : Scout Scan: 279/1/1 Gantry Failed to Stop at Position 90 degrees within 28.000 seconds. (AMC state:7 Cmd:24 CurPos:91.0) Retry Rx to verify probable Axial Drive or Gantry Balance or Brake or Encoder Hardware.

Error: 260107212 Gantry Failed to change to Commanded Velocity of x.x secs/rev within x.000 seconds. Velocity is dependent on what was selected Wed Aug 25 19:08:26 2004 Error: 260107212 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3557 Wed Aug 25 19:08:25 2004 Host : Table Gantry Processor Ermes # : 260107212 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3557 Function : AXIAL CONTROL : Axial Control Scan Type : Axial Scan: 598/1/1 Gantry Failed to change to Commanded Velocity of 2.0 secs/rev within 20.000 seconds. (AMC state:4 Cmd:21

Applies to

Models
BrightSpeed, VCT
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues