GE Healthcare · CTBrightSpeed26011520880582588585388591126011519950030
BrightSpeed HPower, Delight, Velocity or MERI HPower
Symptom
Table will not move up, especially when in full down position. See long description Table will not raise or move up, especially when it's in the full down position and this could be intermittent along with noises noticed by the customer.
Error message
Error: 260115208 Elevation moved to invalid direction See long description for complete error Fri Jan 30 00:21:21 2004 260115208 tgp tgp_subsys mtcb_emc_motion_check.c 249 Fri Jan 30 00:21:21 2004 Host : Table Sub-System Ermes # : 260115208 Exception Class : Soft Severity : Pri/Soft File : mtcb_emc_motion_check.c Line# : 249 Function : No System Function Reported Scan Type : None/Unknown Scan: 9/1/0 Elevation moved to invalid direction. expected target position = 805825 um actual direction = UP last position = 885853 um current position = 885911 um Error: 260115199 Elevation encoder fault See LD for complete error Mon Feb 13 08:20:09 2006 Error: 260115199 Host: tgp Process: tgp_subsys File: mtcb_emc_isrfunc.c Line: 125 Mon Feb 13 08:20:10 2006 Host : Table Sub-System Ermes # : 260115199 Exception Class : Soft Severity : Pri/Soft File : mtcb_emc_isrfunc.c Line# : 125 Function : No System Function Reported Scan Type : None/Unknown Scan: 50030/10/0 Elevation encoder fault.
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2337217 elevation encoder assembly used on HPower Systems. See Long Description for details Part #: 2337217 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ENCODER ASSY, ELEVATION - H POWER FRU: Yes Status: Orderable Parts Offered as: Current: 2337217 Harvested: Not Available Exchange: Not Available Comments: FLS - ENCODER ASSY, ELEVATION - H POWER. ADDED PER CT PCN 220839. MS20544 3/31/03.
- Area
- CT
Full cause & step-by-step resolution