GE Healthcare · MRExcite224932010819348302249323224932122493672249369
Cold Reset Message Occurs During Reboot due to Three Cables Not Connected to APS Board.
Symptom
Cold Reset Message occurs during Reboot. System will not boot. The system will fail to come up and will post a message box asking "Do you want to do a cold reset?"
Error message
2249320 Communication problem between the APS, AGP, and SCP boards. Timeout: apsProbeComplete not received. 1081934830 6 1 Wed Apr 14 09:27:10 2004 2249320 2 CARD diagCoord systemResourceApi.c 525 Task ID: 0x3f1fdf8 Time: 3249 Object: AGP Communication problem between the APS, AGP, and SCP boards. Timeout: apsProbeComplete not received.-> 2249323 Communication problem between the APS, AGP, and SCP boards. Timeout: scpStartConvert not received.-> 1081934830 2 1 Wed Apr 14 09:27:10 2004 2249323 2 CARD diagCoord systemResourceApi.c 644 Task ID: 0xf99df8 Time: 3176 Object: SCP Communication problem between the APS, AGP, and SCP boards. Timeout: scpStartConvert not received.-> 2249321 Communication problem between the APS, AGP, and SCP boards. Timeout: apsConvertComplete not received. 1081934830 8 1 Wed Apr 14 09:27:10 2004 2249321 2 CARD diagCoord systemResourceApi.c 570 Task ID: 0x3f1fdf8 Time: 5254 Object: AGP Communication problem between the APS, AGP, and SCP boards. Timeout: apsConvertComplete not received.-> 2249367 getBoardRevision failed. Board type = 10; status = 0. Wed Apr 14 09:27:10 2004 2249367 4 CARD diagCoord mgdPosixSync.c 476 Task ID: 0xf99df8 Time: 3176 Object: SCP
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.