GE Healthcare · MRExcite3000004224942047275623
HDe TPS Reset Fails Due to VRE Configuration.
Symptom
During Application start-up, some or all of the Image Compute Nodes (ICN) may not successfully initiate and TPS Reset failures will occur See Resolution concept for the complete SMILe released 07/21/2006. Problem: Changing the host IP address or MGD IP address without performing an ICN Reconfiguration leaves the VRE configuration files with information pertaining to the old IP addresses. In this state, during Application start-up, some or all of the Image Compute Nodes (ICN) may not successfully initiate and TPS Reset failures will occur.
Error message
3000004 An exception has occurred in MxJVM Mon Mar 6 05:57:20 2006 Error: 3000004 Host: GH-GEMR Process: LxMxConsumer File: MxJVM Line: 0 An exception has occurred in MxJVM.^C Exception: Mon 03/06/2006 05:57:20.672 AM EST::::Stack Trace Follows: Exception: jav 2249420 Object: SCP A Monarch critical error was detected in task excLogTask (0x0). Error: 2249420 Host: MR01 Process: diagCoord File: excLog.c Line: 1392 Task ID: 0x3dcdef0 Time: 47275623 Object: SCP A Monarch critical error was detected in task excLogTask (0x0). Additional information in /usr/g/service/log/mxExLg-2d15c34
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 14.0 For Hino Excite HDe 14.0 systems only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.