GE Healthcare · MRExcite2252243
can link is interrupted due to GP3 module failure.system hangs intermitently.replace GP3 Module PN
Symptom
HP Computer Hang
Error message
2252243 Network Management Failure detected on the CAN network. A CAN node stopped responding to messages. Failing Node is Master Gradient.-> See Long Description Fri Feb 8 10:42:15 2008 Error: 2252243 Host: SLGIRMA01 Process: NSP File: SCP:CanLinkHWO Line: 993 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: t1memp Coil: 8HRBRAIN Scan: 7544/7 Network Management Failure detected on the CAN network. A CAN node stopped responding to messages. Failing Node is Master Gradient.-> Extended Error : The CAN node indicated stopped responding to watchdog messages from the SCP. Check that the node has power and that the CAN cable is connected. This could also happen if a node was manually powered off.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 2331400-2 Gradient Processor GP3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.