Pular para o conteúdo principal
GE Healthcare · MRSignaExcite222552622255162225540

Low Voltage on the High Voltage SSM Supply.

Symptom

Unable to Scan with Inductive or Direct Drive Bias Fault Unable to Scan with Inductive or Direct Drive Bias Fault. Sat Apr 23 10:21:22 2005 Error: 2225526 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6100 Inductive or Direct Drive fault, or Body TR switch fault due to impedance mismatch, path short Drive bias fault. Possibly a shorted diode in the driver switches, or a short circuit load.

Error message

2225526 Inductive or Direct Drive fault. Drive bias fault 1999 Suite: MROC Host: OC0 Proc: HPC_PORT_ID Error: 2225526 File: PMH.m 1.29 Method: spi_tr_mon Line: 6100 Inductive or Direct Drive fault. Drive bias fault. Possibly a shorted diode in the the driver switches, or a short circuit load. Sat Apr 23 10:21:22 2005 Error: 2225526 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6100 Inductive or Direct Drive fault, or Body TR switch fault due to impedance mismatch, path short. Drive bias fault. Possibly a shorted diode in the the driver switches, or a short circuit load.

2225516 SSM fault, High voltage power supply low. Mon Jun 28 08:02:10 2004 Error: 2225516 Host: pmri2 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6040 SSM fault.^A High voltage power supply low. Also may be a shorted load on dynamic disable or direct drive circuits.

2225540 Dynamic disable fault, Front of Coil short. Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on FRONT of BRM. 2225540 Dynamic disable fault, Front of Coil short Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorte

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
8.3 8.3
Component
SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.