GE Healthcare · MRExcite110041102611002
Illegal Character in the Destination Hostname (space).
Symptom
Cannot Move or Transfer Images on DICOM Network Cannot Move Images on DICOM Network
Error message
11004 Failure of DCM_assoc_action DCM_OPEN_REQ for C-STORE SCU Error: 11004 Host: SRSAMRI Process: dcserver File: dcsclient.c Line: 1087 Failure of DCM_assoc_action: DCM_OPEN_REQ for C-STORE SCU 11026 Failure to find destination in SdCREMOTEHOST_FILE Error: 11026 Host: SRSAMRI Process: dcserver File: dcscommon.c Line: 363 Failure to find destination in SdCREMOTEHOST_FILE 11002 DCM low level error Application error (505) invalid called application title (159) Error: 11002 Host: SRSAMRI Process: dcserver File: dcscommon.c Line: 227 DCM low level error: Application error (505): invalid called application title(159)
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.