Pular para o conteúdo principal
GE Healthcare · MRExcite2262133

Coil Not Upgraded With Coil ID Box.

Symptom

Coil ID wouldn't detect 3, 5 inch or GP flex coils

Error message

2262133 SRI coil validation error: Coil ID command echo not received.Connector: Head T/R Upper Port-> 2262133 RTMR01 NSP SCP:SRI 316 SRI coil validation error: Coil ID command echo not received. Connector: Head T/R Upper Port->

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2383623 1.5T SINGLE CHANNEL RECEIVE ONLY CABLE WITH COIL ID Need Quick Disconnect Box 2383623 for surface coils that connect to top connector of Dog House.
Area
MR

Full cause & step-by-step resolution