GE Healthcare · MRExciteSigna2248016327542244573

Gradient Error 2244573 GP Reporting X Axis SGA Internal Wiring Fault Due to SGA Control Board.

Symptom

Intermittent DW-EPI failures with 2244573 Diffusion scans would intermittently fail to complete with 2244573 "The GP is reporting % Axis SGA Internal Wiring Fault and Internal Power Supply Undervoltage" In this case the failure was always Z axis only. Diffusion scans would intermittently fail to complete with 2244573 "The GP is reporting Z Axis SGA Internal Wiring Fault and Internal Power Supply Undervoltage" In this case the failure was always Z axis only.

Error message

2248016 Warning MR Signal is too high (overrange) Reduce R1 DRF ID 1 FE ID 0 channel 0 2248016 Host: mr waltham Process: NSP File: APS:DRF Line: 563 Warning: MR Signal is too high (overrange); Reduce R1 DRF ID: 1 FE ID: 0 channel: 0 arg1: 32754

2244573 The GP reported a Gradient Amplifier % Axis Internal Wiring Fault or Control Undervoltage Error: 2244573 Host: WMR2 Process: NSP File: SCP:Grad Line: 2587 The GP reported a Gradient Amplifier X Axis Internal Wiring Fault or Control Undervoltage.->

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2250375-4 CONTROL_BOARD ASSEMBLY_SGA 2250375-4 CONTROL_BOARD ASSEMBLY_SGA 2250375-4 Part Desc: CONTROL_BOARD ASSEMBLY_SGA
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.