GE Healthcare · MRExcite226123922275562252239112143041722521332218661
Short Circuit in UPM-ASC Chassis Loading Down +24 VDC CAN Power Supply inside Driver Module.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2261239 ASC NB IF Board was not detected but is configured. See Long Description Fri Jul 15 11:46:26 2005 2261239 lx-mr NSP SCP:CanAutoConfig 2587 ASC NB IF Board was not detected but is configured in MRconfig.cfg. If this board is present, verify power, RF Amp and CAN communication and perform a TPS Reset. If the SSM is present, set usssType=no in MRconfig.cfg and perform a TPS Reset.-> 2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. *** 2252239 Startup Timeout was detected on the Universal Power Monitor 2. 1121430417 3 1 Fri Jul 15 12:26:57 2005 2252239 2 lx-mr NSP SCP:CanAutoConfig 2576 Startup Timeout was detected on the Universal Power Monitor 2.-> EN 168 SR 259 2252239 Startup Timeout was detected on the Master Gradient. Mon May 23 22:49:44 2005 Error: 2252239 Host: signa2 Process: NSP File: SCP:CanAutoConfig Line: 1840 Startup Timeout was detected on the Master Gradient. 2252133 SDO Communication Failure was detected on the CANopen Protocol. Node Detecting Failure = SCP. Node ID = 2. Index = 0x200d. SubIndex = 0x0. Error Code = 0x1001.-> 2252133 FHMR01 NSP SCP:sdo_segment_download 2029 SDO Commu
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2363041-2 UPM-ASC Chassis Assembly Excite HD
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.