GE Healthcare · MRExcite122981202729051
Application Software does not complete the boot process with TPS Download failure.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
9001 Unable to set Connection server AQSserver host D3T 9001Unable to set Connection server AQSserver host D3T 3005 Can't connect with XXX AQSserver. 3005 Can't connect with XXX AQSserver (for message from XXX QRS.12298, code 7008, tag 1202729051) Mon Feb 11 07:24:02 2008 3005 4 MRS2_OC0 netserver netproc.c 536 Can't connect with MRS2_OC0:AQSserver (for message from MRS2_OC0:QRS.12298, code: 7008, tag: 1202729051)
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2272740 ETHERNET HUB SERVICE KIT From Excite HD Parts List
- Area
- MR