SID Issue, use lift orange button to move detector to highest position error 1000014.
Symptom
SID Issue, use lift orange button to move detector to highest position error 1000014. That button is purple now SID Issue, use lift orange button to move detector to highest position error 1000014, That button is purple now 2000 OM Here is the link to the Innova2000 Operator Manuals: Heroic2 LVA FMI http://gein.euro.med.ge.com/tech_publications/docs/doc/heroicpdf/print-copy-enLVAFMI.pdf Heroic2VCP: http://gein.euro.med.ge.com/tech_publications/docs/doc/pdf/2394280-247_r1/om/en/HelpHomePage-en.htm SvcPack: http://gein.euro.med.ge.com/tech_publications/docs/doc/pdf/2405170-247_r2/om/en/HelpHomePage-en.htm They all have the following section: Chapter Innova Positioner Digital Detector Housing Controls Emergency Back-Out Raise (only) the Digital Detector by pressing one of the Emergency Back-Out buttons located on the Digital Detector (See Illustration). Note: This control must be used only: 1. To clear a collision Level 2 (see Section Collision Sensors) by allowing a 2.5 cm additional SID course. 2. To move away the Digital Detector from the patient in case of gantry failure. 3. In case of SID failure. Use this button to bring the Digital Detector to its highest position to still allow X-Ray.
Error message
error Code = 1000014, error Text = SID_SSD out of range, debug Text = SID out of range, recovery Class = IH, major Function /= XRImChainPos System impact: X-ray unavailable. Displayed message : "SID pb, use lift orange button to move detector to highest position". Note: This error is also logged when the positioner is resetted.
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.