Pular para o conteúdo principal
GE Healthcare · MRExcite22445782244575

Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244578 The GP is reporting a Gradient Amplifier % Axis Current Distortion Fault

2244575 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault Error: 2244575 Host: WMR2 Process: NSP File: SCP:Grad Line: 2683 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault.->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
Area
MR

Full cause & step-by-step resolution