GE Healthcare · MRExcite

PM Assist says PM Complete but says Please run SPT in PM Mode Due to Three Passes of SPT in PM Mode are Needed with OPEN…

PM Assist says PM Complete but says Please run SPT in PM Mode Due to Three Passes of SPT in PM Mode are Needed with OPENMR4 M4.

Symptom

PM Assist says PM Complete but says Please run SPT in PM Mode After running LV Shim and SPT in PM mode per documented procedure, PM Assist comes back with PM complete, but says "Please run SPT in PM Mode". If the FE leaves the system this way the customer will get pop-up messages indicating that PM is overdue. Openspeed Excite systems are required to run 2 passes of PM Mode SPT. First selecting Head and then selecting Body. PM check will find the two files and merge them for determining PM completion After running LV Shim and SPT in PM mode per documented procedure, PM Assist comes back with PM complete, but says "Please run SPT in PM Mode". If the FE leaves the system this way the customer will get pop-up messages indicating that PM is overdue. Openspeed Excite systems are required to run 2 passes of PM Mode SPT. First selecting Head and then selecting Body. PM check will find the two files and merge them for determining PM completion

Error message

PM Assist Status Incomplete PM Please RUN SPT Test in PM mode. See long description ***************************************************************** PM Assist Status : PM complete. ***************************************************************** SPT PM FILE NAME : /usr/g/service/data/G4BE2345_PM.SPT ---------------------------------------------------------------- "Z Isocenter - " "Gradcal X - " "Gradcal Y - " "Gradcal Z - " "Head SNR - PASS" "Head FSE Stab. - PASS" "Head GRE Stab. - PASS" "Body SNR - PASS" "Body FSE Stab. - PASS" "Body GRE Stab. - PASS" "Coherent Noise - PASS" ***************************************************************** PM Assist Status : Incomplete PM, Please RUN SPT Test in PM mode. ***************************************************************** LVS PM FILE NAME : /usr/g/service/data/G4BD35RC.LVS ---------------------------------------------------------------- LV Shim : PASS Please Fix the SEVERE Failures (except for Eddy Current or EPIWP tests ) before : 5/2/2006 Last & Next scheduled PM dates ------------------------------- LAST PM DATE : 04/11/2006 Next PM is DUE by : 7/31/2006

Applies to

Models
Excite
Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues