GE Healthcare · X-ray VascularInnova50111
50111:Encoder/pot mismatch: Hardware status 0=L-arm 1=PVT 2=ARC 3=II Lift 50111:Encoder/pot mismatch Hardware…
Encoder/pot mismatch
Symptom
50111:Encoder/pot mismatch: Hardware status 0=L-arm 1=PVT 2=ARC 3=II Lift 50111:Encoder/pot mismatch Hardware status 0=L-arm 1=PVT 2=ARC 3=II 3= Lift failure, dose is minimized
Error message
50111 DO LC Gantry POS SMA: Encoder, potentiometer failure Debug: hw-status=2 pos-error=4abc usr-status=1000 POS ERROR=4ABC,HDW STATUS=0002,USR STATUS=1000 50111 "POS : SMA: Encoder, potentiometer failure, debug Text += pos-error=4ABC, hw-status=0001, usr-status=1000 , recovery Class = DO, major Function /= LC Gantry System impact: No gantry motion. Displayed message : None. 50111 "POS : SMA: ENCODER, POTENTIOMETER FAILURE " DO POS 0 2 HWStatus=0001 User=0000 1000 4ABC 50111 "POS : SMA: Encoder, potentiometer failure, debug Text += pos-error=4ABC, hw-status=0001, usr-status=1000 , recovery Class = DO, major Function /= LC Gantry. System impact: No gantry motion. Displayed message : None.
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Component
- Positioner Gantry Positioner Gantry Positioner Gantry
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.