GE Healthcare · X-ray VascularInnova2800154
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Symptom
Xray not available; Reset system to recover, Call Service is displayed in the Image monitor.
Error message
error Code = 2800154, error Text = Runtime failure of Viewer0, debug Text += Process runtime fatal error, recovery Class = IH, major Function /= XRSubSysMgrDL System impact: Exam failure. Displayed message : "Xray not available - Reset system; if problem persists, call Service".
Applies to
- Models
- Innova
- Product
- X Innova 4100 X Do not use---------4100 systems 4100 systems
- Software
- Unity 3D M3B kit Unity 3D M3B kit, coming from FMI. DL Release : DL_UNITY_3.16.1 Ghost DL : 21.15 DL : DL_Unity_3.16.1 DL Atlas Core : 4.2-8_3 RTAC Atlas Core : 4.2-8_4 IDC : 5.18 Calypso : 4.24 Calypso Uptime : Unity_CLYUPT_3.8 Positioner : 1.4.3U_8 Jedi : P6.34RC1
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.