GE Healthcare · MRExciteSigna

High level of B0 ghosting caused by poor short cross term compensation.

Symptom

ACR tests indicate a high level of B0 ghosting on fMRI sequences. See test results in long description ACR tests indicate a high level of B0 ghosting on fMRI sequences.b0 image results as measured by ept b0 image were high 4-5 % but within system specifications. ACR test suggested typical results should be 1-2% on their test.

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR