Pular para o conteúdo principal
GE Healthcare · MRExcite2247160511762247159511742247157

TPS Reset Fails Due to RRF Exciter. 3T HD Excite 12.0. 2247157 The Master Exciter in the RRF chassis (rear, slot 7) was

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2247160 RRF Fiber Cable Link Not Ready See Long Description Fri Jan 25 09:23:14 2008 Error: 2247160 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 272 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51176 Object: APS Test Task RRF FIBER CABLE LINK NOT READY The fiber link between the RRF and MGD chassis is not in a ready state after hardware initialization. This will prevent the MGD from controlling the RRF chassis. Try TPS_reset. Run RRF diagnostics.-> Extended Error: The MGD does not like the signal coming over the fiber cable connected from the IRF_IO board in the back of the MGD chassis to the DIF board in slot 10 in the front of the RRF chassis. Possible cause(s): RRF-DIF Board, GigaLink Fiber Cable, IRF I/O Board, IRF Board

2247159 Sine Wave Table on RIF DIF1 Board Did Not Load Correctly See Long Description Fri Jan 25 09:23:14 2008 Error: 2247159 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 412 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51174 Object: APS Test Task The Sine Wave Table on the DIF board in RRF chassis (front, slot 10) did not load correctly. This will prevent proper Exciter output. Try TPS_rese

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Component
2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues