Pular para o conteúdo principal
GE Healthcare · MRExcite2224666

Cannot Advance Table to Scan Position Due to Clutch was loose causing it to slip when moving the table small increments.

Symptom

Cannot Advance Cradle to Scan Position See Long Description Cradle will not move into bore to landmark, longitudinal motor works but cradle is stuck on patient transport.

Error message

2224666 Auto Prescan failed The table is not at scan plane Wed Oct 19 18:25:22 2005 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, The table is not at scan plane.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues