GE Healthcare · X-ray VascularInnovaAdvantx30068
IN2K, Advantx, 41B4 Tube casing water flow not detected.
Symptom
IN2K, Advantx, 41B4 Tube casing water flow not detected. Call service and continue is displayed on the console during an exam IN2K, Advantx, 41B4 Tube casing water flow not detected. Call service and continue is displayed on the console during an exam
Error message
IN2K, Advantx, 41B4, 42FB, 30068:Tube casing water flow not detected IN2K, Advantx, 41B4, 42FB, 30068:Tube casing water flow not detected: This error does not inhibit Xrays. If the water flow issue is not address, the XRT could overheat enough to open the >70 degree C switch on the XRT, the error 42FB is then generated and time has to be allowed for the XRT to cool. Error 30068 is generated on an Innova 2000 system. IN2K, Advantx 41B4, 42FB, 30068: Tube casing water flow not detected: This error does not inhibit Xrays. If the water flow issue is not address, the XRT could overheat enough to open the >70 degree C switch on the XRT, the error 42FB is then generated and time has to be allowed for the XRT to cool. Error 30068 is generated on an Innova 2000 system.
Applies to
- Models
- Innova, Advantx
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems
- Component
- Gantry C-arc coupling, Rotation Joint pn 2184070-2 LC ROTATING JOINT 2184070-2 at the center of the C-arc inside the cable hose Part #: 2184070-2 Description: PR ROTATING GASKET FRU: Yes Status: Orderable Comments: FLS - RP ROTATING JOINT, REPLACES 2184070 WHEN STOCK IS AT ZERO.PER SPCN JLG 99/12/16-01. STG 12/28/99 QUICK DISCONNECT 2125878 LC ROTATING JOINT 2184070-2 at the center of the C-arc inside the cable hose Part #: 2184070-2 Description: PR ROTATING GASKET FRU: Yes Status: Orderable Comments: FLS - RP ROTATING JOINT, REPLACES 2184070 WHEN STOCK IS AT ZERO.PER SPCN JLG 99/12/16-01. STG 12/28/99 QUICK DISCONNECT 2125878
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.