Preventive Maintenance (PM) Schedule F Requires Cardiac Gating Cable Test Due to an Error in Documentation.
Symptom
Preventive Maintenance PM Schedule F Requires Cardiac Gating Cable Test. But It Does Not Exist See Long Description An FE called the OnLine Center to report that he could not find a Cardiac Gating Cable Test in PM Safety Checks in HDx Service Methods documentation (Direction 5155500, Rev# 5.2, Signa HDx Service Methods) appropriate and valid for this product type. The PM Schedule F requires this test to be performed, and includes documentation to perform it in the linked Magnet Room Maintenance document URL: http://3.87.118.28/optec3/common/hdx/content/640042.htm . However this document is written only for 5x and LX systems , it doesn't pertain to the HDx product and therefore can't be followed when performing the PM on an HDx system. I checked the PM Schedule F document and confirmed what the FE had told me. I then checked elsewhere in HDx Service Methods in an attempt to find a valid Cardiac Gating Cable Test for this product to no avail. iTrak 13150405 submitted.
Applies to
- Models
- Excite, Signa
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.