Preventive Maintenance (PM) Schedule F Requires Cardiac Gating Cable Test Due to an Error in Documentation.

Symptom

Preventive Maintenance PM Schedule F Requires Cardiac Gating Cable Test. But It Does Not Exist See Long Description An FE called the OnLine Center to report that he could not find a Cardiac Gating Cable Test in PM Safety Checks in HDx Service Methods documentation (Direction 5155500, Rev# 5.2, Signa HDx Service Methods) appropriate and valid for this product type. The PM Schedule F requires this test to be performed, and includes documentation to perform it in the linked Magnet Room Maintenance document URL: http://3.87.118.28/optec3/common/hdx/content/640042.htm . However this document is written only for 5x and LX systems , it doesn't pertain to the HDx product and therefore can't be followed when performing the PM on an HDx system. I checked the PM Schedule F document and confirmed what the FE had told me. I then checked elsewhere in HDx Service Methods in an attempt to find a valid Cardiac Gating Cable Test for this product to no avail. iTrak 13150405 submitted.

Applies to

Models
Excite, Signa
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR