GE Healthcare · X-ray VascularInnovaAdvantx400008360010540000803997698

Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.

Symptom

Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76

Error message

4000083 Arcnet chip is not ready to send a packet 4000083 Arcnet chip is not ready to send a packet. Will cause error 600105 Recommended Action: Check the 48V Martek Run RTAC subsystem//PCHost//Gitane_ARCNET_COM diagnostics. verify the power distribution in the Positioner (the Martek modules) check the ARCNET fiber from the RTAC to the positioner

4000080 GEB Drive failed to Xmit the packet ARCNET Communication 4000080 GEB Drive failed to Xmit the packet AR XRTitan logTitanError Component(s) ARCNET Communication 1 0 GebDriver local error code = 2 , error number = 3997698 Tif.cpp

IN2K: Ping Test from Digital Leader to Titan Kernal Fails Ping Test from Digital Leader to Titan Kernal Fails. The Ping from the Digital Leader (DL) to the Titan fails. Attempt to ping the Titan Kernal from the DL. Open a command window and Ping 165.100.10.3. If The system reports "request timed out", The Ping to the Titan Kernal has failed

Applies to

Models
Innova, Advantx
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Area
X-ray Vascular

Full cause & step-by-step resolution