Pular para o conteúdo principal
GE Healthcare · X-ray VascularAdvantxInnova30053212011882

Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.

Symptom

Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76

Error message

30053, error Text = POWER UNIT : KV UNBALANCE FAULT, debug Text = HWStatus=0039 User=0000 0000 42E9, recovery Class = DO, major Function /= ECM System impact: None. 1 tube spit that can be heard by customer. Displayed message : None. ADTX error 42E9. need to have 30053 as a syn psdb cant find it. 212011882 02/03/2011 20:47 GMT

Applies to

Models
Advantx, Innova
Product
Advantx E Biplane Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Advantx, LFX Advantx product line LFX Advantx product line Innova 2000 Innova 2000 systems Innova 2000 systems Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems
Component
Advantx-E VAMP Advantx-E VAMP Advantx-E VAMP
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.