GE Healthcare · X-ray VascularInnova200207
The system will not complete a bolus chase.
Symptom
The system will not complete a bolus chase. Tube collision sensor error is displayed on the the monitor.
Error message
200207 PrepOn requested while the SPIN or Bolus 200207 Software Debug information DO XRAcqSync control Software_Problem 1 3 PrepOn requested while the SPIN_SAFE_TEST inhibit is still active for opMode 6 XRPAtlasLogger.cpp 156 200207 PrepOn requested while the MOTION CONTROL ERROR inhibit is still active for opMode 20 200207 Software Debug information DO XRAcqSync control Software_Problem 1 3 PrepOn requested while the SPIN_SAFE_TEST inhibit is still active for opMode 6 XRPAtlasLogger.cpp 156. This error is logged when the customer attempts to make an acquisition during the Spin Test Blous, 200207 PrepOn requested while the MOTION CONTROL ERROR inhibit is still active for opMode 20
Applies to
- Models
- Innova
- Product
- X Innova 4100 X Do not use---------4100 systems 4100 systems
- Software
- Bolus M4+ DL release : DL_LBYBOLUSM4_10.1.050805 Ghost DL-2 : NA Ghost DL-3 : 18.8 Ghost DL-5 : 20.6 DL : DL_LBYBOLUSM4_10.1.050805 DL Atlas Core : 3.1.R RTAC Atlas Core : 3.1.R IDC : 4.5.5 Calypso : 2.21 Calypso Uptime : BOLUS1.24 Positioner : 1.2.24B_IS6.6 Jedi : B6.30BFC
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.