GE Healthcare · MRExcite222458940627962224713
GAIN STAGE ON RF AMP NOT ADJUSTING PROPERLY.
Symptom
Auto Prescan is slow and getting low signal on Manual Prescan ISSUES WITH RF AMP. TG FLIP ANGLE VERY HIGH ON BODY SCANS. SYSTEM PRESCANS LONGER THAN USUAL ISSUES WITH RF AMP. TG FLIP ANGLE VERY HIGH ON BODY SCANS. SYSTEM PRESCANS LONGER THAN USUAL.
Error message
2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan. Note: On later systems, this same error code has been slightly reworded to read as follows: Thu Jun 12 11:06:12 2008 Error: 2224589 Host: MENEMR Process: NSP File: TPS:aps_exec Line: 4062796 TG reached the high limit. TG has been set to 190. Check patient positioning and if desired try Auto Prescan again or use Manual Prescan. 2224713 No successful FTG check in X iterations APS1 will run automatically. Error: 2224713 Host: DMHMRI Process: NSP File: APS:aps_exec Line: 1 No successful FTG check in 2 iterations. APS1 will run automatically.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.