Unable to Scan With HV Supply Low, Front of Coil Short & Rear of Coil Short Always Logged.

Symptom

Unable to Scan With High Voltage Power Supply Low, Front of Coil Short & Rear of Coil Short Always Logged Customer is unable to scan. The three error messages shown in the Long Description are always logged. 2249147 oahhgeexmrmr NSP SCP:../TrDriverMdsHWO.cpp 1255 SSM fault. High voltage power supply low. Also may be a shorted load on dynamic disable or direct drive circuits. 2249159 oahhgeexmrmr NSP SCP:../TrDriverMdsHWO.cpp 1378 Dynamic disable fault, Front of Coil short Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on FRONT of BRM. 2249161 oahhgeexmrmr NSP SCP:../TrDriverMdsHWO.cpp 1394 Dynamic disable fault, Rear of Coil short Dynamic Disable bias fault on back drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on BACK of BRM.

Error message

2249147 High Voltage Power Supply Low. Also may be a shorted load on dynamic disable or direct drive circuits. Tue Mar 30 13:50:47 2004 Error: 2249147 Host: CCRC Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1255 SSM fault.^A High voltage power supply low. Also may be a shorted load on dynamic disable or direct drive circuits.

2249159 Dynamic disable fault, Front of Coil short. Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on FRONT of BRM. 2249159 lx-mr NSP SCP:../TrDriverMdsHWO.cpp 1683 Dynamic disable fault, Front of Coil short Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on FRONT of BRM

2249161 Dynamic disable fault Rear of Coil short 2249161 Host: pomr Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1699 Dynamic disable fault, Rear of Coil short^A Dynamic Disable bias fault on back drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on BACK of BRM.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.