Missing ECC Coeficient Files Due to Grafidy3 Not Run.

Symptom

Error Messages During Each Scan With No Apparent Effect There seems to be no impact on system performance other than the error log quickly growing very large.

Error message

2252735 AGP could not find the ECC file /export/home/signa/caldir/b0_ecc_coeff.dat.WHOLE to download the coefficients. ECC coefficients will not be loaded to the WARP Please check to see if the ECC file is present 2252735 AGP could not find the ECC file /export/home/signa/caldir/b0_ecc_coeff.dat.WHOLE to download the coefficients. ECC coefficients will not be loaded to the WARP

2261245 SCP Advisory Message. SCP could not open the ECC coefficient file /export/home/signa/caldir/grad_ecc_coeff.dat.ZOOM. Please do restore info or run Grafidy to create the file.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution