Patient comfort trips due to faulty SRI.

Symptom

Will Not Scan With Patient Comfort Sensor Trip. See Long Description Patient comfort sensor trip; cannot scan due to bore sensor indicating temperature is too hot for patient safety. A message "Patient Comfort Sensor Trip" appears in the message box on the left side of the scan page at the start of autoprescan.

Error message

2248655 Hardware Failure: the bore temperature reading is invalid. Error: 2248655 Host: LFHMRI Process: NSP File: SCP:BoreTempSriHWO Line: 770 Hardware Failure: the bore temperature reading is invalid. It is reading too high. Check for an open in the sensor, sensor cable, or connector, or a problem with the amplifier board on the SRI.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
5115910 SRI2 Scan Room Interface II
Area
MR