GE Healthcare · MRExcite13091904

No Automated SNR Test Available in HDx for 4 Channel CTL Coil.

Symptom

No Automated SNR Test Available in HDx and HDxT for 4 Channel CTL Coil or other coils See long descrption. Previous versions of MR service software have included a tool called CTLSNR and/or Multi coil SNR Tool with which an automated SNR test could be run on the four channel CTL coil. In 14.0, these tools were replaced with the Multi-coil Quality Assurance (MCQA) Tool. This new tool ONLY works with coils that support Coil ID. The problem is that when we upgrade a site to 14.0 (HDx), or beyond, we are often left with the older four channel CTL coil but no way to test it for SNR. iTrak 13091904

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution